Chacaltaya GAW Station (LFA - IIF - UMSA )
  • Home
  • About Us
  • Instruments & Data
    • Instruments, data & metadata
    • Aethalometer
    • Automatic Weather Stations
    • CO analyzer
    • CO2 and CH4 analyzers
    • High Volume Air Sampler
    • Hygroscopic Tandem Differential Mobility Analyzer
    • Integrating Nephelometer
    • LIDAR
    • Mobility Particle Sizer Spectrometer
    • Multi-Angle Absorption Photometer
    • Neutral Cluster & Air Ion Spectrometer
    • Ozone Analyzer
    • Passive Atmospheric Sampling
    • Passive Atmospheric Sampling, semi-automated
    • Quadrupole Aerosol Chemical Speciation Monitor
    • SO2 Analyzer
    • Sun photometer
    • Water isotopic composition
  • Staff
    • Current staff LFA (Bolivia)
    • Previous staff LFA (Bolivia)
    • International Consortium
    • Students
  • News
  • Publications
  • Visit us
  • Contact
  • Links
  • GCOS-GRUAN Project
  • Latin American and Caribbean Aerosol Measurements School 2015
    • Motivation and Objective
    • Course description
    • Program
    • Lecturers
    • Bios Lecturers
    • Who should attend?
    • Parallel session for stakeholders
    • Application procedure and registration
    • Financial support
    • Calendar
  • Current staff LFA (Bolivia)
Surface ozone analyzer

Since 2012, surface ozone concentration is measured at Chacaltaya with a Thermo Scientific 49i Ozone Analyzer.

The instrument is a UV photometric analyzer consisting of a dual-cell for differential measurements. As O3 molecules absorb UV light at a wavelength of 254 nm, the degree to which the UV light is absorbed is directly related to the ozone concentration (Beer-Lambert Law). The response time of this instrument is of 20 seconds, and at Chacaltaya data are stored with a one-minute resolution.

Automatic zero and span checks were performed daily until 2017, but due to an internal failure, zero checks started being performed manually every 2 weeks since 2018.

The instrument was calibrated against a traceable standard in 2017 at the WMO regional training center  (Argentina).

Contact person: Isabel Moreno

Picture
Ozone analyser Thermo Sci. 49i